8.2.5 Setup

[0--] Interface Block

You can access the parameters of the instruments’s serial interface(s) in parameter section [0--].
The data output modes can be used only once per instrument except for continuous data output.

[01-] Serial Interface 1

This sub-block includes the parameters of the 1

st

serial interface.

[010 X ] Data Format

0 : No data transfer

1 : Continuous data output

2 : not defined

3 : Host mode

[011 X] Baud Rate

0 : 1 200 Baud

1 : 2 400 Baud

2 : 4 800 Baud

3 : 9 600 Baud

4 : 19 200 Baud

5 : 38 400 Baud

6 : 57 600 Baud

[012 X] Handshake

0 : No Handshake

1 : Xon/Xoff

2 : Hardware

[013 XX] Address

The address range is 1 to 99. If you enter 0, the indicator will operate without an address.

[014 X] Data Length and Parity

0 : 8 bit, no parity

1 : 7 bit, odd parity

2 : 7 bit, even parity

[02-] Serial Interface 2

The parameters of the 2

nd

serial interface.

[020 X ] Data Format

0 : No data transfer

1 : Continuous data output

2 : not defined

3 : Host mode

4 : Modbus RTU

[021 X] Baud Rate

0 : 1 200 Baud

1 : 2 400 Baud

2 : 4 800 Baud

0 : 1 200 Baud

4 : 19 200 Baud

5 : 38 400 Baud

6 : 57 600 Baud

[022 X] Handshake

0 : No Handshake

1 : Xon/Xoff

[023 XX] Address

The address range is 1 to 99. If you enter 0, the indicator will operate without an address.

[024 X] Serial Mode

0 : Interface operates as RS232C (param. [012] ≠ 2)

1 : Interface operates as RS485

[025 X] Data Length and Parity

0 : 8 bit, no parity

1 : 7 bit, odd parity

2 : 7 bit, even parity

Note:

The parameter [025] must be set to “0” for Modbus RTU output.

[04-] Weight Data Transfer

The serial interface between the weighing board and the PC104 board is standard serial interface at baud rate
9600. You can adjust only following features:

[045 ] Minimum Print

[XXXXXX]

If the weight is less than the value entered here, the data will not be printed

[046

X] Print Control

0 : Printing with key

1 : Auto print

2 : Print interlock

Explanation:

If set to “Auto print”, the data will automatically be printed when the value is stable and

higher than the minimum print value. The weight value should fall below the minimum print value to
reprint. If this parameter is set to “Print interlock”, after printing the weight must change for reprint.

FT-16(D) Technical Manual, Rev. 1.21 December 2008

Page 20 of 32