8 communication, 1 overview about communication, 2 serial interface – Flintec FT-16D V.1.21 User Manual

Page 18: 1 standard serial interface connector (rs232c), Communication, Overview about communication, Serial interface, Standard serial interface connector (rs232c), Ommunication

Advertising
background image

8 C

OMMUNICATION

8.1 Overview about Communication

Application Hardware

channel

Operation mode

of the interface

Continous output
mode;
see Chap. 8.2.3

Data export to PC (non-approved);
Remote display;
Remote control with ASCII entries
(Z, T, C)*

Option: Serial Interface 1

(RS232C; see Chap. 8.2.1 & 8.2.5)

OR

Option: Serial Interface 1 & 2

(RS232C/RS485; see Chap. 8.2.2 & 8.2.5)

Host mode;
see Chap. 8.2.4 &
Appendix 2

Process control
(PLC, DCS, SCADA...)

Option: Serial Interface 2 (RS485; see Chap. 8.2.2 & 8.2.5)

Modbus RTU;
see Appendix 3

Connection of a 2nd scale (e.g. FT-
11(D)) for display and data
acquistion; Serial printer

Serial standard interface of the PC104 board

(RS232C; setup via application software)

* Z = Zero; T = Tare; C = Clear; P = Print

8.2 Serial

Interface

8.2.1 Standard Serial Interface Connector (RS232C)

The PC104 board has a standard serial interface for connecting to peripherals. The configuration is done by the
application software.

Standard Serial interface

Baudrate

1200, 2400, 4800, 9600, 19 200, 38 400 or 57 600 baud

Data format

8 bit no parity OR 7 bit even parity OR 7 bit odd parity

Start / Stop bit

1 start bit and 1stop bit


The connection of the standard RS232C of the instrument should be made as given in the table below.

Definition

Pin no. (D-Sub, 9-pin, male)

RS232C; typical PC allocation

TXD 3

3

RXD 2

2

GND 5

5

Shield Connector

body


You have to connect each TXD (output) with RXD (input) at the other end.

8.2.2 Optional Serial Interface Connector (RS232C, RS485)

Additional serial interfaces can be added to the weighing board in addition to the standard RS232C port.

The optional serial interfaces are suitable for bi-directional communication. If you transmit ASCII codes of P, Z,
T or C letters to the serial interface, the indicator will behave like the related keys have been pressed.
If optional serial interfaces are installed, only continuous data output can be programmed for more than one
interface (see chapter 8.2.5 for setup details).

1

st

Serial interface

Optional RS232C; Parameter group [01-]

2

nd

Serial interface

Optional RS232C / RS485 (field changeable); Parameter group [02-]


Up to 32 instruments can be connected with enabled RS485 port.

FT-16(D) Technical Manual, Rev. 1.21 December 2008

Page 18 of 32