7.5. OPTIONAL

ACCESSORIES

IMPORTANT
The installation of optional accessories must be carried out by authorised SMEG
technical staff.

7.5.1. PERISTALTIC PUMPS P1 AND P3 (ACCESSORY “SMEG ADU”)


During first start-up or whenever the detergent can runs empty, it is necessary to activate peristaltic pumps
loading using programme 8.

The detergent recommended for use with P1 is smeg Deterliquid D2.

The pump dosage can be adjusted by following the procedure described in paragraph "SETUP MODE
DESCRIPTION".

Similarly for P3, an optional peristaltic pump, available only for model GW1060, to be used to introduce extra
detergent.

GW2045
The optional pump P1 must be mounted on the rear cross section, in the free space next to the pump P2. On
the back of the pre cut shape, which must be removed to install P1, the terminal boards for the pump’s
electrical connection can be found.
There is a metal plate fixed with Philips screws on the rear, above the pumps. The terminal board is fixed to
the inside of the plate, which allows electrical connection of the optional level sensors.

fig. 13 - Rear of GW2045, position of pumps and detail of plate that holds electrical connection terminal board.

GW1060

The optional peristaltic pump must be mounted on the rear, in the free space next to the pump P2. Upper
position for P1, lower position for P3.
The metal rear of the machine is divided into two sections, looking from behind: Left and right. The two
sections can be easily divided (they are fixed by Philips screws and inserts with ¼ turn and can be opened
with a flat screwdriver).
The right side is the one holding the peristaltic pumps. To mount the optional pump it is necessary to remove
the precut shapes on the framework. The electrical connection terminal boards for the pumps and optional
level sensors can be found on the back of the shapes.