Recommended disinfection protocol – Soniclean UW3000 User Manual
Page 22

RECOMMENDED DISINFECTION PROTOCOL
SONICLEAN UW-3000 IRRIGATED ULTRASONIC
CLEANER
It is Soniclean’s protocol to disinfect every Washer supplied for trial at the end
of every trial and immediately prior to commencing the next trial. Soniclean or
their distributors/agents will ensure this is undertaken unless the trial
organisation prefers to conduct the disinfection themselves.
The protocol to be followed:
1.Drain Washer and turn off at power point.
2. Apply protective gloves (disposable non sterile latex gloves are
suitable).
3.Remove tray; thoroughly wipe down all visible surfaces of the Washer
including the lid, tank, all outside surfaces, hoses and attachments
with disinfectant solution
1
.
4.Remove detergent bottle and replace with bottle containig slightly
more than required amount of disinfectant
1.2
5.Turn Washer on at power point whilst holding down the program key
(this will allow you to re-program the parameters).
6.Change soap % parameter to required amount of disinfectant. Ensure
irrigate function on.
7.Press “clean”. The Washer will now add appropriate amount of
disinfectant to tank and in doing so will disinfect the internal
plumbing fittings as well as Washer tank.
UW-3000 Washer Operating Manual
Page 22 of 24
© Soniclean Pty Ltd
Version 2.0
April 2009
KEY TO PROGRAMMING DISPLAY
PARAMETER
DISPLAY
MACHINE
MIN
MAX
FACTORY
SETTINGS
Clean time
Ct
mins
1
99
10
Ultrasonic power level
PO
1
3
3
Rinse time
rt
mins
0.1
9.9
0.4
No. of cleans between drains
Cn
0
99
10
Fill temperature
F
O
0
2
0
Rinse temperature
r
O
0
3
3
Detergent concentration
SP
percentage
0.0
5.0
0.5
Start up degassing time
G1
mins
0=Off
99
5
Test time
tt
sec
0=Off
99
30