8. Once clean is complete, disconnect disinfectant bottle and replace

with bottle filled with clean water.

9.Press “shift” and “refill” keys, hold for three seconds. This will cause

Washer to empty of water and disinfectant and then to fill with water.
This will rinse disinfectant out of the unit.

10.Repeat steps 5 and reprogram soap parameter to factory setting (i.e.

1.2)

11.Dry all external surfaces and pack unit ready for transport.
___________________________________________________________________

Notes:

1.

Disinfectant solutions

: The recommendation is to use a medical grade

disinfectant, TGA approved, that is recommended for use in hospital settings
for disinfecting horizontal and vertical working surfaces.

Soniclean currently uses a product called LCC, CSSD (supplied by Dominant
Chemicals). For disinfecting the Washer it is recommended that you dilute
LCC, 1:35. That is:

1.1To disinfect surfaces you will need to add 6mls of disinfectant to

200mls of water and then use this solution to wipe the surfaces
with.

1.2To disinfect internal plumbing fittings, set the soap % parameter

to 5. (This will ensure
1200mls of LCC is added to the 45itre tank).