Operating instructions, . how to operate, Start process – Soniclean UW3000 User Manual
Page 7: Confirm parameters, Load the machine

4. Operating Instructions
4.1. . How to Operate:
4.1.1. Start Process:
i)
Turn on at power point.
ii)
The machine will display 8.8.8.8, then SCx.x
iii)The machine will automatically drain, flush, drain the flush, fill with water
and detergent and perform a de-gassing cycle.
iv)When the degassing cycle is complete, the machine is ready for use.
v)During the above cycle, the machine with display the following;
Display
Function
8.8.8.8
Internal program
SCX.X
Internal program process
Drn
Draining the water.
Drnf
Tank is flushing. (drain is open)
Drn
Tank is draining the flush
FILL
Tank is filling with water and
detergent is automatically added
at the programmed quantity.
(0.5%)
De-gas
The solution is degassing. This
takes approximately 10 minutes.
----
The machine is ready to be
loaded with instruments.
vii)
It is recommended that you perform a foil test daily to check the
performance of your Washer (Refer Section 5, Foil Test)
4.1.2. Confirm Parameters:
To check or change the settings or parameters of your Washer:
i)
Press the “SHIFT” key and “PROG” key together and hold for 3 seconds.
ii)
A beep will sound when command has been recognized.
iii)
The machine will display the “Clean Time” or “ct”
iv)
Scroll through the parameters using the “Enter” keys
v)
The parameters may be changed as required. (Refer Section 6, Program
Mode Section)
4.1.3. Load the Machine:
i)
Instruments may be loaded into tray with the tray located on a bench
top.
ii)
Gently lay the instruments on the bottom of the tray .
iii)
Lower the tray into the cleaning bath.
UW-3000 Washer Operating Manual
Page 7 of 24
© Soniclean Pty Ltd
Version 2.0
April 2009