Hazardous location use (FDO

®

70x IQ H models)

Rated sensor models

The following sensor models are rated for hazardous location use:

 FDO

®

700 IQ H (YSI-Order No. 207065Y)

 FDO

®

701 IQ H (YSI-Order No. 207066Y)

Hazardous location rated sensors are identified by the name plate with
the rating details engraved on the sensor shaft:

Name plate of a Hazloc

sensor (example)

DANGER

Explosion hazard.
Only hazardous location rated sensor models
("Hazloc Sensor") must be used in hazardous
locations. Read the name plate on the sensor shaft
and verify that the sensor is rated for your specific
application.

DANGER

Explosion hazard.
The associated apparatus MIQ/BB1 (YSI part no.
207001Y) has to be used for connection. Refer to the
following documents for proper installation:
Control drawing, YSI document no. 28123

(see MIQ/BB1 operating manual)

MIQ/BB1 operating manual
All applicable electrical code regulations.

DANGER

Explosion hazard.
This product does not meet the requirements of the
directive 94/9/EC (ATEX).

######

#########

###############

####################

#################

######

######

FDO 700 IQ H

28123