9 - 2

ba76014e03

05/2014

Indexes

FDO

®

70x IQ ...

EC833x

Sensor could not be calibrated,
sensor blocked for measurement
Cause: instable signal
* Check temperature adjustment
* Check calibration conditions (see operating manual)
* Repeat calibration

EC933x

Calibration error, measurement disabled
Cause: Sensor cap is missing, leaky, depleted, or defective
* Clean sensor and space between cap and sensor according to
op.instructions
* Screw on sensor cap securely
* Repeat calibration
* Replace sensor cap

EI133x

Operational voltage too low
* Check installation and cable lengths, Follow installation instructions
* Power supply module(s) overloaded, add power supply module(s)
* Check terminal and module connections
* Defective components, replace components

EI233x

Operational voltage too low, no operation possible
* Check installation and cable lengths, Follow installation instructions
* Power supply module(s) overloaded, add power supply module(s)
* Check terminal and module connections
* Defective components,
replace components

ES133x

Component hardware defective
* Contact service

Message code

Message text