4 instrument safety, 1 general instrument safety, 2 hazardous location ratings (h models) – YSI IQ SensorNet FDO Sensor User Manual

Page 37: 5 electrical data, Instrument safety -3 7.4.1, General instrument safety -3, Hazardous location ratings (h models) -3, Electrical data -3

Advertising
background image

FDO

®

70x IQ ...

Technical data

ba76014e03

05/2014

7 - 3

7.4

Instrument safety

7.4.1

General instrument safety

7.4.2

Hazardous location ratings (H models)

Applicable directives

and standards

In addition to the standards listed in chapter 7.4.1, the FDO

®

70x IQ H

conforms to the following directives and standards:

 ANSI/ISA-12.12.01

 CAN/CSA C22.2 # 213

The FDO

®

70x IQ H is a non-incendive electrical equipment for use in

Class I Division 2 hazardous locations.

Hazardous location

rating

7.5

Electrical data

Applicable norms

– UL 61010-1

– CAN/CSA C22.2 # 61010.1

FDO

®

700 IQ H

FDO

®

701 IQ H

Class I Division 2
Groups A, B, C, D T6
Class I Zone 2
Group IIC T6

Associated apparatus

MIQ/BB1, YSI part no. 207001Y

Control Drawing

YSI document no. 28123

Rated voltage
(non-H models)

24 V DC
via the IQ

S

ENSOR

N

ET

(for more

details, see chapter

T

ECHNICAL

DATA

of the IQ

S

ENSOR

N

ET

sys-

tem operating manual)

Voltage range
(H models)

14 V DC ... 24 V DC via the MIQ/
BB1 barrier box (=associated
apparatus)

Power consumption

0.7 W

Protective class

III

Advertising