9 indexes, 1 explanation of the messages, 1 error messages – YSI IQ SensorNet FDO Sensor User Manual

Page 43: Indexes -1, Explanation of the messages -1 9.1.1, Error messages -1, 9indexes

Advertising
background image

FDO

®

70x IQ ...

Indexes

ba76014e03

05/2014

9 - 1

9

Indexes

9.1

Explanation of the messages

This chapter contains a list of all the message codes and related
message texts that can occur in the log book of the IQ

S

ENSOR

N

ET

system for the FDO

®

70x IQ ... sensor.

9.1.1

Error messages

Information on the contents and structure of the log book,
and how to call it up, is given in the

L

OG

BOOK

chapter of the

IQ

S

ENSOR

N

ET

system operating manual.

The last three digits of the message code identify the
source of the message:

 334 = SC FDO 700 (sensor cap SC-FDO

®

700)

 335 = SC FDO 701 (sensor cap SC-FDO

®

701)

 531 = FDO700IQ (sensor basis / component class,

adapter ADA)

Message code

Message text

EA133x

Meas. range exceeded or undercut
* Check process
* Select other meas. range

EA233x

Sensor temperature too high!
* Check process and application

EA333x

Sensor temperature too low!
* Check process and application

EAP33x

Measurement interfered
* SensCheck: Sensor cap is missing, leaky, depleted, or defective
* Clean sensor and space between cap and sensor according to
op.instructions
* Screw on sensor cap securely
* Replace sensor cap

EAP531

Measurement interfered
* SensCheck: Sensor cap is missing, leaky, depleted, or defective
* Clean sensor and space between cap and sensor according to
op.instructions
* Screw on sensor cap securely
* Replace sensor cap

Advertising