Measuring / Operation
ba76014e03
05/2014
4 - 7
4.2.5
Reactivating previous calibration data
The FDO
®
70x IQ ... enables you to reactivate the last valid user cali-
bration or the factory calibration. Thus you can immediately go on
measuring if a calibration procedure failed or you suspect that the cali-
bration conditions were not optimally met.
Reactivating calibration
data
Reactivating old calibration data is only a temporary solu-
tion. Please take into account that the sensor might thus
possibly provide incorrect measured values. Please make
sure the sensor works correctly by carrying out a function
check and/or recalibration.
1
Open the setting table (see section 3.4).
2
In the Calibration menu item, select the User calibration or Fac-
tory calibration setting and then quit the setting table with Save
and quit.