3 measuring the temperature, 2 orp calibration, Measuring the temperature – YSI MultiLab 4010-2 User Manual
Page 46: Orp calibration

ORP voltage
MultiLab 4010-2
46
ba76142e03
08/2013
6.1.3 Measuring the temperature
For reproducible ORP measurements, it is essential to measure the tempera-
ture of the test sample.
When operating a sensor without integrated temperature sensor, you first have
to measure and enter the temperature of the sample.
The measuring instrument recognizes whether a suitable sensor is connected
and automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature measuring
mode:
6.2
ORP calibration
Temperature
sensor
Resolution of
the temp. dis-
play
Temp. measurement
Yes
0.1 °C
Automatic with temperature sensor
-
1 °C
Manual
ORP electrodes are not calibrated. You can, however, check ORP
electrodes by measuring the ORP of a test solution and comparing
the value with the nominal value.