YSI MultiLab 4010-2 User Manual
Page 68

Ion concentration
MultiLab 4010-2
68
ba76142e03
08/2013
12.
Add the standard solution to the sample.
13.
Using <> <> and <ENTER>, enter the values for the volume of the
standard solution (Std. volume) and concentration of the standard solu-
tion (Std. conc.).
14.
Select continue with <> <> and start the measurement with
<ENTER>. The measurement result is displayed when the measure-
ment is completed.
15.
If necessary, start measuring further samples with <ENTER>.
Repeat steps 6 - 14 for all samples.
16.
Terminate the measuring method with <M>.
A safety query appears.
17.
Select yes with <> <>.
18.
Confirm yes with <ENTER>.
Measurement with the selected method is completed.
i
Add standard!
Std. volume
100.0 ml
Std. conc.
1.0 mg/l
continue
Blank value addition
01.08.2013 08:00
01.08.2013 08:00
Blank value addition