YSI MultiLab 4010-2 User Manual
Page 51

MultiLab 4010-2
Ion concentration
ba76142e03
08/2013
51
Continuing with two-
point calibration
9.
Immerse the electrode in standard solution 1.
10.
When calibrating without temperature sensor:
Measure the temperature of the standard solution using a thermom-
eter.
Use <F4>/[
⇅ ] to select the setting of the temperature.
Use <> <> to set the temperature.
Use <F4>/[
⇅ ] to select the setting of the concentration.
11.
Set the concentration of the standard solution with <> <> and
press <ENTER>.
The standard solution is measured.
The measured value is checked for stability (AutoRead).
12.
Wait for the end of the AutoRead measurement or accept the calibra-
tion value with <ENTER>.
The calibration display for the next standard solution appears.
01.08.2013 08:00
01.08.2013 08:00
Standard
13.
Thoroughly rinse the electrode with distilled water.
14.
Immerse the electrode in standard solution 2.