YSI MultiLab 4010-2 User Manual
Page 63

MultiLab 4010-2
Ion concentration
ba76142e03
08/2013
63
10.
Using <> <> and <ENTER>, enter the values for the volume of the
standard solution (Std. volume), the concentration of the standard solu-
tion (Std. conc.) and the volume of the ISA/TISAB solution (ISA/TISAB
vol.).
11.
Select continue with <> <> and start the measurement with
<ENTER>.
When the measurement is finished an entry window appears.
12.
Add the sample to the standard solution.
13.
Using <> <> and <ENTER>, enter the value for the volume of the
sample (Sample volume).
14.
Select continue with <> <> and start the measurement with
<ENTER>.
The measurement result is displayed when the measurement is com-
pleted.
15.
If necessary, start measuring further samples with <ENTER>.
Repeat steps 6 - 12 for all samples.
16.
Terminate the measuring method with <M>.
A safety query appears.
17.
Select yes with <> <>.
18.
Confirm yes with <ENTER>.
Measurement with the selected method is completed.
i
Add sample!
Sample volume
1.0 ml
continue
Sample addition
01.08.2013 08:00
01.08.2013 08:00
Sample addition