Accu-Chek Aviva Expert User Manual

Page 103

Advertising
background image

95

4

NOTE

• If data are not available to display the trend graph, the No Data Available screen is displayed.

• Trend graphs do not include control results.
• If (time change icon) is shown at the bottom of the Trend Graph screen, the time and date of one or more

data points in the trend graph may not match the time stamp in My Data because the meter clock has been

changed. All times in the trend graph show the data relative to the current meter time.

For example:

1. You take a bG test and the value is stored in the meter at 9:00 am.

2. The meter time is changed ahead 1 hour.

3. The bG data on the trend graph is shown at 10:00 am.

• Trend graph data is cleared if bolus advice data is cleared.

• On the right side of the graph, the Maximum Bolus Amount Indicator is a horizontal line with the

corresponding value shown above it. The purpose of this indicator is to scale the top portion of the graph for

the Bolus Value blue bars. The Maximum Bolus Amount Indicator values available on the meter are 1, 5, 15,

30, and 60 U. The indicator displayed is based on the single largest bolus insulin amount delivered for the

time range chosen. For example, if the largest bolus delivered during the chosen time range is 8 U, then the

meter will scale the upper graph to be between 0 and 15 U.

• On the right side of the graph, the Maximum Carbs Amount Indicator is a horizontal line with the

corresponding value shown below it. The purpose of this indicator is to scale the bottom portion of the graph

for the Carb Value dark orange bars. The Maximum Carbs Amount Indicator values available on the meter are

30, 60, 120, 180, and 240 g, or the equivalent scale for BE, KE, or CC. The indicator displayed is based on

the single largest carbs amount for the time range chosen. For example, if the largest carbs amount for the

chosen time range is 86 g, then the meter will scale the lower graph to be between 0 and 120 g.

• Select Table to change the display to the Table View. The time range remains the same.

• The displayed insulin only includes bolus values.