256

10.4 Product Safety Information

WARNING

Choking hazard. Small parts. Keep away from children under the age of 3 years.

Electromagnetic Compatibility
This meter meets the electromagnetic immunity requirements as per EN ISO 15197 Annex A. The chosen
basis for electrostatic discharge immunity testing was basic standard IEC 61000-4-2. In addition, it meets
the electromagnetic emissions requirements as per EN 61326. Its electromagnetic emission is thus low.
Interference from other electrically driven equipment is not to be anticipated.

Performance Analysis
The ACCU-CHEK Aviva Expert system (ACCU-CHEK Aviva Expert meter with ACCU-CHEK Aviva Plus test strips)
is calibrated with venous blood containing various levels of glucose. The reference values are obtained using
the hexokinase method. For method comparison, the results were compared with results obtained using the
hexokinase method with deproteinization (automatic analyzer). The hexokinase method is traceable to an
NIST standard.