Out-of-Range Control Test Results

If your control test result is out of range, check this list to help to solve the problem:

Troubleshooting Checks

Actions

1. Were either the test strips or control solutions

expired?

Discard the test strips or control solutions if either is
past the Use By date. If the control solution was
opened more than 3 months ago, discard it. Repeat
the control test with an unexpired test strip and an
unexpired control solution.

2. Did you wipe the tip of the control solution bottle

before use?

Wipe the tip of the bottle with a tissue. Repeat the
control test with a new test strip and a fresh drop of
control solution.

3. Were the caps on the test strip container and the

control solution bottle always closed tightly?

Replace the test strips or control solutions if you
think either was uncapped for some time and repeat
the control test.

4. Was the test strip used immediately after it was

removed from the test strip container?

Repeat the control test with a new test strip and a
fresh drop of control solution.

5. Were the test strips and control solutions stored in

a cool, dry place?

Repeat the control test with a properly stored test
strip and control solution.