Accu-Chek Aviva Expert User Manual

Page 232

Advertising
background image

224

4.

j

Carefully wipe

around the test

strip slot and other openings.
Make sure that no liquid enters
any opening.

5.

j

A separate Super Sani-Cloth

should be used for cleaning
and disinfection. For
disinfecting the meter, get a
new cloth and repeat steps 3
and 4, making sure that the
surface stays wet for
2 minutes.

NOTE

Roche has tested the approved product for a total of 208 cleaning and disinfecting cycles. This supports
cleaning and disinfecting the meter once per week for a 3 year time period and allows for 52 extra
cleaning and disinfecting cycles that might occur (due to blood contamination, for example) over the
lifetime of the meter. This equates to approximately 1 additional cleaning and disinfecting cycle per month
over the 3 year use life of the meter.

6

j

Wash hands thoroughly with

soap and water.

52123_07006322001.indb 224

10/18/13 2:32 PM

Advertising