Accu-Chek Aviva Expert User Manual

Page 255

Advertising
background image

247

9

Display Shows Possible Cause(s)

Possible Solution(s)

Invalid Hypo Values

j

The hypo warning limit value

must be less than all of your
target ranges in the time
block settings.

j

Reset the hypo warning limit or revise the target

ranges in time blocks and re-enter the hypo
warning limit.

j

Enter a hypo warning limit that is below the target

ranges of your time blocks. See Chapter 5,
Changing Meter Settings.

Invalid Record
Time/Date

j

The time/date entered is

invalid (Add Data entries
cannot be set to occur in the
future).

j

Re-enter time/date.

Invalid Target
Range

j

The lower target range value

is above the upper target
range value.

j

Reset the target range values.

j

Enter the correct lower target range value and

upper target range value. See Chapter 5, Changing
Meter Settings.

j

The range selected conflicts

with your hyper and/or hypo
warning limit settings.

j

Reset range or revise warning limit settings and

re-enter range.

j

Enter the correct lower target range value and

upper target range value. See Chapter 5, Changing
Meter Settings.