Display Shows Possible Cause(s)

Possible Solution(s)

E-57 Electronic
Error

j

An electronic error has

occurred or, in rare cases, a
used test strip was removed
and reinserted.

j

Turn off the meter and remove the batteries. Wait at

least 20 seconds prior to reinserting the batteries.
Turn on the meter and perform a blood glucose or
control test. If the problem persists, contact
ACCU-CHEK Customer Care Service Center at
1-800-688-4578.

E-58 Temp. Error

j

The temperature is above or

below the proper range for
the meter.

j

Move the meter to an area within the proper

temperature range indicated for test strip use in the
test strip package insert. Wait 5 minutes before
turning on the meter. Repeat the test. Do not
artificially heat or cool the meter.

E-59 Battery
Empty

j

The batteries are extremely

low.

j

Insert new batteries. See Chapter 8, Maintenance.

E-60 Time/Date
Error

j

The time and date settings

may be incorrect.

j

You have changed the

batteries.

j

Ensure the time and date are correct and adjust, if

necessary. See Chapter 5, Changing Meter
Settings.