SPECTRUM Series

ENMET Corporation

15

Table 2: Spectrum Calibration Voltage and Countdown Times

N

OTE

:

The following are values used for production calibration. Not all gases are available for field calibration;

contact

ENMET

customer service personnel.

Gas

Calibration Gas

* Calibration Cup

Calibration Voltage

Countdown Time

Ammonia

25ppm

*

03620-011

0.67

180 sec

Arsine

(2)

0.5ppm PH

3

= 0.38ppm

03620-018

0.10

120 sec

Carbon Monoxide

100ppm

*

03620-011

1.70

120 sec

Chlorine

(2)

5ppm

03620-011

1.40

180 sec

Fluorine

(2)

3ppm Cl

2

= 7.5ppm

03620-011

1.80

240 sec

Hydrazine

0.3ppm

TBD

0.91

240 sec

Hydrogen

800ppm

*

03620-011

2.20

180 sec

Hydrogen Chloride

(2)

10ppm

03620-018

2.10

180 sec

Hydrogen Cyanide

10ppm

*

03620-011

2.10

120 sec

Hydrogen Fluoride

(1)

5ppm

TBD

1.40

240 sec

Hydrogen Sulfide

20ppm

*

03620-011

0.46

120 sec

Nitric Oxide

25ppm

TBD

0.67

90 sec

Nitrogen Dioxide

5ppm

*

03620-011

1.30

120 sec

Oxygen

20.9%(air)

*

03620-011

2.80

120 sec

Ozone

(1)

0.3ppm

03620-018

0.75

240 sec

Phosgene

(1)

0.66ppm

03620-018

0.18

240 sec

Phosphine

(2)

0.5ppm

*

03620-011

0.13

120 sec

Silane

5ppm

TBD

0.54

120 sec

Sulfur Dioxide

10ppm

TBD

2.10

90 sec

N

OTE

:

All specifications stated in this manual may change without notice.

TBD

– To Be Determined

(1) Require use of Sample Draw System. See Figure 6.
(2) Recommended use of Sample Draw System, for enhanced sensitivity and speed of response. See Figure 6.

*

Initial Calibration Cup 02552-008

For Sensor Replacement

Sampling Adapter / Calibration Cup

03620-011

For Standard Calibration

Initial Calibration Cup 02552-023

For Sensor Replacement

Sampling Adapter/ Calibration Cup

03620-018

Example of Sample Draw System
with Sampling Adapter 03620-018

Figure 6: Identification of Calibration Adapters & Sample Draw System

To Sensor

Motorized Sample Pump

03700-029

MOTORIZED SAMPLE

PUMP

OUT

IN

CHARGE

JACK

ALARM

POWER

Sample OUT
To Sensor

Ambient Air Sample IN

To Sensor

To Sensor

To Sensor