ENMET SPECTRUM User Manual

Page 6

Advertising
background image

ENMET Corporation

SPECTRUM Series

2

1.2 Turn Instrument ON

Turn the instrument ON, by pressing and holding the

POWER / BACKLIGHT

pushbutton for two seconds. In

uncontaminated air, for most instruments the display should read 0000 within ten seconds of turn-on. For an oxygen
Spectrum, the display reads near 20.9%.

N

OTE

:

Instruments using biased sensor, this time is extended to 4 minutes, stabilization may take as long as 1 hour.

See Table 1.

1.2.1 Verify

The

SPECTRUM

is calibrated prior to shipment. However, if there is access to a source of the target gas, such as a

calibration kit, expose the sensor to the gas, and observe that the instrument is responsive.

1.2.2 Acknowledge Alarm

If the concentration of the target gas is greater than the alarm set point, the instrument indicates an alarm condition.
Acknowledge the alarm by pressing and releasing the right hand pushbutton, SELECT; this silences the audio alarm
for four minutes unless the concentration of the target gas is greater than the upper alarm limit. See Table 1 for a list
of alarm set points and upper alarm limits for various target gases.

1.2.3 Remove Gas

Remove the source of the target gas. After the display reads zero or close to it, turn the instrument OFF, by pressing
and holding the

POWER / BACKLIGHT

pushbutton for approximately three seconds. The display flashes "OFF" and

then goes blank after the pushbutton is released.

1.2.4 Contact

ENMET

If the instrument doesn't operate as described, contact

ENMET

customer service personnel immediately.

Figure 1: SECTRUM Features

Display

Sensor

Front View

Visual Alarm

Audio Alarm

Side View

Cover

Access to Battery

Pushbutton

Switches

Top View

Advertising