ENMET SPECTRUM User Manual

Page 24

Advertising
background image

ENMET Corporation

SPECTRUM Series

20

SPECTRUM Series for Nitric Oxide (3E 100 sensor)

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

chlorine

5

nitrogen dioxide

100

hydrogen

1000

carbon monoxide

1000

carbon dioxide

10000

saturated HC, alcohols

abundant

sulfur dioxide

50

SPECTRUM Series for Nitrogen Dioxide (3E sensor)

The following gases cause a reading:

Interferant
Gas/Vapor

Interferant Concentration
in ppm

Instrument Reading
in ppm

chlorine

1

3

ozone

1

0.7

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

hydrogen chloride

5

nitrogen oxide

100

hydrogen

1000

carbon monoxide

1000

carbon dioxide

10000

saturated HC, alcohols

abundant

sulfur dioxide

50

ammonia

30

SPECTRUM Series for Hydrazine (2E sensor)

The following gases cause a reading:

Interferant
Gas/Vapor

Interferant Concentration
in ppm

Instrument Reading
in ppm

ammonia

200

0,4

methyl hydrazine

1

0,72

1.1 dimethyl hydrazine

1

0,45

chlorine

3

–0,02

hydrogen sulfice

20

0,1

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

ethanol

1000

Isopropanol

450

hydrogen peroxide

10

hydrogen chloride

7

methanol

1200

hydrogen

2000

carbon monoxide

1000

carbon dioxide

5000

unsaturated HC (ethylene)

1%