Appendix B: Interference Gases

Appendix B: Interference Gases

Appendix B: Interference Gases

Appendix B: Interference Gases

Interference Table for Selected Spectrum Sensors

SPECTRUM Series for AMMONIA (3E-100 sensor)

The following gases cause a reading:

Interferant
Gas/Vapor

Interferant
Concentration in ppm

Instrument
Reading in ppm

ethanol

1000

1

isopropanol

450

1

hydrogen sulfide

14

10

methanol

1200

3

hydrogen

1000

80

carbon monoxide

300

100

sulfur dioxide

25

-21

nitrogen dioxide

50

25

hydrogen cyanide

10

-18

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

chlorine

5

unsaturated hc (ethylene)

1.0 %

saturated hydrocarbons

abundant

SPECTRUM Series for CARBON MONOXIDE (3E sensor)

The following gases cause a reading:

Interferant
Gas/Vapor

Interferant Concentration
in ppm

Instrument Reading
in ppm

hydrogen

1000

450

nitrogen oxide

100

25

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

ammonia

100

chlorine

5

hydrogen cyanide

10

ethylene

2.0 % *

carbon dioxide

5,000

methane

10,000

sulfur dioxide

10 *

hydrogen sulfide

10 *

nitrogen dioxide

10 *

isopropanol

1,025 *

gasoline vapor

saturated *

*with onboard filter; continuous high level exposure may reduce the filter efficiency