ENMET SPECTRUM User Manual

Page 26

Advertising
background image

ENMET Corporation

SPECTRUM Series

22

SPECTRUM Series for Sulphur Dioxide(4S sensor)

The following gases cause a reading:

Interferant
Gas/Vapor

Interferant Concentration
in ppm

Instrument Reading
in ppm

carbon monoxide

300

<3

hydrogen sulphide

15

0

nitric oxide

35

0

nitrogen dioxide

5

≈-5

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

chlorine

1

hydrogen

200

hydrogen cyanide

10

hydrogen chloride

5

ethylene

100

SPECTRUM Series for Arsine (2E sensor)

The following gases cause a reading:

Interferant
Gas/Vapor

Interferant Concentration
in ppm

Instrument Reading
in ppm

phosphine

100 ppb

110

chlorine

5

-400

hydrogen cyanide

10

100

ammonia

100

10

diborane

100

35

silane

1

100

germane

1.1

100

hydrogene selenide

50 ppb

5

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

hydrogen sulfide

1*

sulfur dioxide

2

hydrochloric acid

5

carbon monoxide

300

carbon dioxide

5000

nitrogen

100 %

hydrocarbons

% range

hydrogen

1000

*with onboard filter; continuous high level exposure may reduce the filter efficiency