Sense[1..n]:correction:offset [:magnitude – EXFO FTB -5240S/5240BP Optical Spectrum Analyzer User Manual

Page 553

Advertising
background image

SCPI Command Reference

OSA

547

Product-Specific Commands—Description

:SENSe[1..n]:CORRection:OFFSet

[:MAGNitude]

Description

This command sets the power offset that is

added to every point measured by the

instrument.

At *RST, this value is set to 0.0 dB.

Syntax

:SENSe[1..n]:CORRection:OFFSet[:MAGNitude]

<wsp><Offset[<wsp>DB|W/W|PCT]>|MAXi

mum|MINimum|DEFault

Parameter(s)

Offset:
The program data syntax for <Offset> is defined

as a <numeric_value> element followed by an

optional <SUFFIX PROGRAM DATA> element.

The allowed <SUFFIX PROGRAM DATA>

elements are: DB|W/W|PCT. The <Offset>

special forms MINimum, MAXimum and DEFault

are accepted on input.

MINimum allows to set the instrument to the

smallest supported value.

MAXimum allows to set the instrument to the

greatest supported value.

Advertising