EXFO FTB -5240S/5240BP Optical Spectrum Analyzer User Manual
Page 623

SCPI Command Reference
OSA
617
Examples on Using the SCPI Commands
TRAC:FEED:CONT "TRC1", NEXT
Reactivate WDM analysis trace feed control for
another "one-shot" refresh.
TRAC:FEED:CONT? "TRC1"
Wait for trace refresh to be done. Poll trace feed
until the returned value is NEVER.
<Add commands to retrieve analysis
results>
See Retrieving Analysis Results (WDM) on
…
Continue intermediate results queries as
necessary.
TRAC:FEED:CONT "TRC1", ALW
Ready to stop acquisition, set the WDM analysis
trace feed to ALWays to make sure that the last
acquired trace is analyzed and updated results
are available once the acquisition loop is
completed.
INIT:CONT OFF
Stop sweep acquisition loop.
STAT:OPER:BIT8:COND?
Wait for the measurement to be completed.
Poll bit 8 until the returned value is 0.
<Add commands to retrieve analysis
results>
See Retrieving Analysis Results (WDM) on
Command Sequence
Comments