Performing a custom inband acquisition (wdm) – EXFO FTB -5240S/5240BP Optical Spectrum Analyzer User Manual
Page 620

SCPI Command Reference
614
FTB-5240S/S-P/BP
Examples on Using the SCPI Commands
Performing a Custom InBand Acquisition (WDM)
STAT:OPER:BIT8:COND?
Wait for the acquisition to be completed. Poll
bit 8 until the returned value is 0.
<Add commands to retrieve analysis
results>
See Retrieving Analysis Results (WDM) on
Command Sequence
Comments
<Add commands to configure Wdm
analysis parameters>
See Configuring the Analysis Setup for the Next
Acquisition Sequence (WDM) on page 603.
CALC:WDM:CHAN:AUTO:NOIS:AUTO
OFF
Optional: if the default channel is active then
set auto noise to disabled.
CALC:WDM:CHAN:AUTO:NOIS:TYPE
INB
Optional: if the default channel is active then
set the specific InBand noise measurement
type.
CALC:WDM:CHAN:SEL "C_001"
CALC:WDM:CHAN:NOIS:AUTO OFF
Set the selected channel auto noise to
disabled.
CALC:WDM:CHAN:NOIS:TYPE INB
Set the selected channel specific InBand noise
measurement type.
CALC:WDM:CHAN:SEL "C_002"
CALC:WDM:CHAN:NOIS:AUTO OFF
Set the selected channel auto noise to
disabled.
CALC:WDM:CHAN:NOIS:TYPE INBN
Set selected channel specific InBand noise
measurement type
CALC:WDM:CHAN:SEL "C_003"
CALC:WDM:CHAN:NOIS:AUTO OFF
Set the selected channel auto noise to
disabled.
Command Sequence
Comments