Initiate:continuous – EXFO PM-1100 Optical Power Meter User Manual
Page 90

Remote Control Commands
80
PM-1100
Product-Specific Commands—Description
INITiate:CONTinuous
Description This command starts or stops continuous mode. When
continuous mode is activated, the latest value is continuously
stored in the initialized memory location so it can be fetched at
any time with the FETC[:SCAL]:POW:DC? query. Values can be
power or loss measurements, depending on the current
measurement mode of the PM-1100.
Syntax
INIT:CONT<space><boolean>
Parameters The <boolean> value can be
“0” to stop continuous mode
“1” to start continuous mode
Example
INIT:CONT 1
Note
Use the INIT:STOR command to initialize a memory location.
See also
FETC:SCAL:POW:DC?, INIT:CONT?, INIT:IMM, INIT:STOR, and
READ:SCAL:POW:DC?
INITiate:CONTinuous?
Description This query returns a value indicating whether continuous mode
is activated.
Syntax
INIT:CONT?
Response
“0” if continuous mode is stopped
“1” if continuous mode is started
Example
INIT:CONT?
See also
FETC:SCAL:POW:DC?, INIT:CONT, INIT:IMM, and
READ:SCAL:POW:DC?