Source[1..n]:wavelength:sweep: repeat – EXFO IQS-2600B Tunable Laser Source for IQS Platforms User Manual
Page 105

SCPI Command Reference
Tunable Laser Source
99
Product-Specific Commands—Description
:SOURce[1..n]:WAVelength:SWEep:
REPeat
Description
This command is used to select the sweep
looping mode. The module can perform the
sweep continuously or execute the sweep
according to the number of cycles specified.
At *RST, the repeat mode depends on the
number of cycles.
Syntax
:SOURce[1..n]:WAVelength:SWEep:REPeat<wsp
>CONTinuous|NCONtinuous
Parameter(s)
Repeat:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
CONTinuous|NCONtinuous.
The parameter is indicating whether or not the
sweep will be continuous:
CONTinuous, the sweep is continuous.
NCONtinuous, the sweep will be executed
according to the number of cycles specified.
Example(s)
SOUR:WAV:SWE:REP CONT
Notes
This function cannot be performed while the
module is sweeping.
See Also
SOURce[1..n]:WAVelength:SWEep:REPeat?
SOURce[1..n]:WAVelength:SWEep:CYCLe