Source[1..n]:wavelength:sweep: time – EXFO IQS-2600B Tunable Laser Source for IQS Platforms User Manual
Page 121

SCPI Command Reference
Tunable Laser Source
115
Product-Specific Commands—Description
:SOURce[1..n]:WAVelength:SWEep:
TIME?
Description
This query returns the continuous sweep
duration. This duration represents the time
required to reach the stop point from start point
at a specified speed.
At *RST, the value is a function of the start point,
the stop point and the sweep speed.
Syntax
:SOURce[1..n]:WAVelength:SWEep:TIME?
Parameter(s)
None
Response Syntax
<ExecutionTime>
Response(s)
ExecutionTime:
The response data syntax for <ExecutionTime>
is defined as a <NR3 NUMERIC RESPONSE
DATA> element.
The <ExecutionTime> response is a value
representing the duration currently set for the
continuous sweep in second.
Example(s)
SOUR:WAV:SWE:START 1550.000nm
SOUR:WAV:SWE:STOP 1555.000nm
SOUR:WAV:SWE:SPE 5.0nm
SOUR:WAV:SWE:TIME? returns 1.000000E+000
See Also
SOURce[1..n]:WAVelength:SWEep:SPEed?
SOURce[1..n]:WAVelength:SWEep:STARt
SOURce[1..n]:WAVelength:SWEep:STOP