Source[1..n]:wavelength:sweep:stop – EXFO IQS-2600B Tunable Laser Source for IQS Platforms User Manual

Page 111

Advertising
background image

SCPI Command Reference

Tunable Laser Source

105

Product-Specific Commands—Description

:SOURce[1..n]:WAVelength:SWEep:STOP

Description

This command allows you to specify a sweep

stop wavelength setting.

At *RST, the stop wavelength is 1570 nm.

Syntax

:SOURce[1..n]:WAVelength:SWEep:STOP<wsp

><StopWavelength[<wsp>M|HZ]>|MAXimum

|MINimum|DEFault

Parameter(s)

StopWavelength:
The program data syntax for <StopWavelength>

is defined as a <numeric_value> element

followed by an optional <SUFFIX PROGRAM

DATA> element. The allowed <SUFFIX

PROGRAM DATA> elements are: M|HZ. The

<StopWavelength> special forms MINimum,

MAXimum and DEFault are accepted on input.

MINimum allows to set the instrument to the

smallest supported value.

MAXimum allows to set the instrument to the

greatest supported value.

DEFault allows the instrument to select a value

for the <StopWavelength> parameter.

The <StopWavelength> parameter is the desired

stop spectrum value for the sweep. If you do not

specify a unit, the device will use the current

spectral units (meter or Hz). To select the current

spectral units, use the UNIT[1..n]:SPECtrum

command.

Advertising