Source[1..n]:power[:level] [:immediate][:amplitude – EXFO IQS-2600B Tunable Laser Source for IQS Platforms User Manual
Page 84

SCPI Command Reference
78
IQS-2600B
Product-Specific Commands—Description
:SOURce[1..n]:POWer[:LEVel]
[:IMMediate][:AMPLitude]
Description
This command sets the source output power
level. This value can be set even if the source is
off.
At *RST, the power level is -8.0 dBm.
Syntax
:SOURce[1..n]:POWer[:LEVel][:IMMediate][:AM
PLitude]<wsp><OutputPower[<wsp>DBM]>|
MAXimum|MINimum|DEFault
Parameter(s)
OutputPower:
The program data syntax for <OutputPower> is
defined as a <numeric_value> element
followed by an optional <SUFFIX PROGRAM
DATA> element. The allowed <SUFFIX
PROGRAM DATA> element is DBM. The
<OutputPower> special forms MINimum,
MAXimum and DEFault are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
DEFault allows the instrument to select a value
for the <OutputPower> parameter.
This parameter is the new source output power
value in dBm.