Source[1..n]:wavelength:sweep: dwell – EXFO IQS-2600B Tunable Laser Source for IQS Platforms User Manual
Page 128

SCPI Command Reference
122
IQS-2600B
Product-Specific Commands—Description
:SOURce[1..n]:WAVelength:SWEep:
DWELl?
Description
This query returns the time interval between
each step of the Step-by-Step sweep mode.
At *RST, the time interval between each step is 2
seconds.
Syntax
:SOURce[1..n]:WAVelength:SWEep:DWELl?[<w
sp>MAXimum|MINimum|DEFault]
Parameter(s)
Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MAXimum|MINimum|DEFault.
MINimum is used to retrieve the instrument's
smallest supported value.
MAXimum is used to retrieve the instrument's
greatest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax
<DwellTime>