Chapter 3: Getting Started

Product Service 1-800-522-7658

9

Connecting the Electrical Supply Source

to the Enclosure

115V Models

Simply connect the 115V power cord supplied to the IEC electrical supply plug on
the back of the enclosure.

230V Models

The same procedure applies for the 230V except it is shipped without a plug.
Install the appropriate plug for your electrical specifications per local codes.

Do not use any detachable power cord that is not adequately rated
for the unit.

Ne pas utliser un fil électrique amovible qui n’est pas du tension
nominale de l’appareil.

Validating the Fume Adsorber Face

Velocity

To determine the actual face velocity at the sash opening, airflow velocity readings
are taken. The “average face velocity” is achieved by taking readings in two rows
across the enclosure with the readings 6" from the ends and evenly spaced every
12". The average velocity across the opening of the enclosure is 35 fpm and should
be validated. Do not use the enclosure with an average velocity of less than 25
fpm. Consult your safety officer.

NOTE: Face velocity profiles and smoke testing should be
performed at least monthly
.


!

115V

Maximum

circuit load

is 2 amps.

230V

Maximum

circuit load

is 1 amp.

!