Monitoring airflow and changing the carbon filters (if equipped) is the primary
maintenance required.

Review this chapter on maintenance for the following:

1. Routine Maintenance.

2. Determination of when to replace carbon filters and how to replace.

3. Calculating carbon filter life.

Routine Maintenance Schedule

Weekly

Wipe down the interior and exterior surfaces of the enclosure with a mild

soap, depending upon the usage of the unit, and allow to dry.

Operate the exhaust system, noting the airflow velocity through the

enclosure using a source of visible smoke.


Monthly (or more often as required)

Determine the actual face velocity through the sash opening of the

enclosure where the average reading should be at the specified velocity of
35±10 fpm. (Use calibrated thermal anemometer or other approved
apparatus).

While the enclosure is filled with the contaminant, test filter condition on

carbon filters using the appropriate gas detector tube at intervals of 20% of
the total estimated time. There is a 5/16" diameter hole on the upper right
side for inserting the detector tube. The exception to the 20%

Advertising