Chapter 5: Using The Station

Product Service 1-800-522-7658

12

Dusting in the Station

Start-up

Turn the fan switch to “Fan.”

Check the condition of the prefilter and note if the filter system light

is on. Ensure that the perforated work surface is not obstructed.

Wear a long sleeved lab coat with knit cuffs and over-the-cuff rubber

gloves. Use protective eyewear. Wear a protective mask if
appropriate.

Wipe-Down

Wipe down the perforated work surface with 70% ethanol, or a

suitable cleaner, and allow it to dry.

CAUTION: Do not use bleach on the stainless steel work surface. Use
disinfectants compatible with stainless steel.

Dusting Evidence

Position evidence during dusting so that air flows around all surfaces

of it, and directly into the perforated work surface. Do not overload
the station. Do not place objects directly on top of the perforated work
surface in a manner that would obstruct the downflow.

Do not obstruct the perforated work surface or the exhaust air

discharge, on top of the station.

Shutdown

Turn off the station. Wipe down the perforated work surface with

70% ethanol, or a suitable cleaner, and allow it to dry.

Remove the perforated work surface. Powder can be cleaned out of

the pan, or the pan can be removed from the filter module and
dumped in a waste container.

Remove the filter cover. Inspect the prefilter, and replace it if

necessary.

Reinstall the filter cover, work surface pan and perforated work

surface. Dispose of rubber gloves appropriately, and have lab coat
laundered properly.