Refer to the following table if the station fails to operate properly. If the
suggested corrective actions do not solve the problem, contact Labconco for
additional assistance.

PROBLEM

CAUSE

CORRECTIVE ACTION

Station blower won’t
turn on

Unit not plugged into
outlet

Plug the station into appropriate
electrical service.

Check connection to the top of station.

Circuit breaker
tripped

Reset circuit breaker.

Filter System Light
turns on

Blockage of the
prefilter

Check the prefilter; replace if needed.
Ensure that there is no obstruction of
the exhaust on top of the station.

HEPA filter is
blocked

If replacing prefilter does not help,
replace the HEPA filter.