Refer to the following table if the station fails to operate properly. If the
suggested corrective actions do not solve the problem, contact Labconco for
additional assistance.
PROBLEM
CAUSE
CORRECTIVE ACTION
Station blower won’t
turn on
Unit not plugged into
outlet
Plug the station into appropriate
electrical service.
Check connection to the top of station.
Circuit breaker
tripped
Reset circuit breaker.
Filter System Light
turns on
Blockage of the
prefilter
Check the prefilter; replace if needed.
Ensure that there is no obstruction of
the exhaust on top of the station.
HEPA filter is
blocked
If replacing prefilter does not help,
replace the HEPA filter.