Chapter 2: Prerequisites

Product Service 1-800-522-7658

3

Location Requirements

Note: The station should be located on a flat, level bench or work surface for
proper operation.

Exhaust Requirements

As stated earlier, the exhaust discharge of the station, located on the upper
left side on the back of the station should never be obstructed. The station is
designed to have its back surface pushed against a wall, but the top of the
station should have no obstructions within 6 inches of the exhaust discharge.

Electrical Requirements

The Downdraft Station models have the following electrical requirements:

Model #

Requirements

3410000-01

115 VAC, 60 Hz, 1 Amp

3410020-25

230 VAC, 50/60 Hz, 0.5 Amp

Note: An outlet with a circuit breaker rated at 10 amps for 115 volt
models (5 amps for 230 volt models) should be located as close as possible
to the right rear side of the station. Power cords supplied with the
station are 3 meters (10 feet) long.

Table 2-1