Safety precautions – Labconco Protecto Series Evidence Drying Cabinets 3390001 User Manual

Page 15

Advertising
background image

Chapter 4: Performance Features and Safety Precautions

Product Service 1-800-522-7658

11

Safety Precautions

For all models:
NEVER expose ANY EVIDENCE, plastic or coated materials to ultraviolet
(UV) radiation. Never bypass the UV safety interlock that only allows the
UV light to work when the door is closed. UV exposure to the skin is
hazardous and can result in serious eye or skin injury.
Ensure that the cabinet is connected to electrical service in accordance with
local and national electrical codes. Failure to do so may create a fire or
electrical hazard. Do not remove or service any electrical components
without first disconnecting the cabinet from electrical service.

NEVER use this cabinet as a chemical storage cabinet.
When handling evidence, or working in the cabinet, always wear appropriate
personnel protective equipment.
NEVER attempt to dry evidence containing flammable gases or solvents in
the cabinet. An open flame should NOT be used in the cabinet.

HEPA filters only remove particulate matter. Operations generating volatile
toxic chemicals are prohibited. The media of the filter is fragile and should
not be touched. Avoid puncturing the HEPA filter during installation or
normal operation. If you suspect that a HEPA filter has been damaged, DO
NOT use the cabinet; contact Labconco at 800-821-5525 or 816-333-8811 for
further information.

The HEPA filters in the cabinet will gradually accumulate airborne
particulate matter from the items placed in the cabinet. The rate of
accumulation will depend upon the cleanliness of the air, operating time and
the nature of items being dried in the cabinet.

Do not operate the cabinet without all of the appropriate filters in place. If an
odor is detected coming out of the granular activated carbon, it needs to be
replaced.
The weight limit for the fold down shelves is 15 lbs (6.8 kg), and the floor of
the cabinet and lower storage area is 20 lbs (9.1 kg).

When surface disinfecting the cabinet:

• Always wear appropriate personnel protective equipment.

• NEVER use high pressure water to clean the liner; this practice could

generate biohazardous aerosols.

• NEVER spray disinfectants into the blower intake at the top of the

liner; this may damage the blower.

• Avoid splashing the disinfecting solution on skin or clothing.

• Ensure adequate room ventilation.

• Carefully follow the disinfectant’s safety instructions.

Advertising