Chapter 5: Using The Cabinet

Product Service 1-800-522-7658

15

Drying in the Cabinet


Start-up

• Turn off UV light and turn the UV Security switch to “OFF”. Remove

the key, and store it in a safe place.

• Turn the switch to “Fan”.

• Check the condition of the prefilters. If your protocol requires the use

of evidence prefilters, ensure a new set of both door and cabinet
prefilters has been properly installed, according to the instructions in
Chapter 6: Maintaining Your Cabinet. Note if the Filter System light
is on.

• Wash hands and arms thoroughly with germicidal soap.

• Wear a long sleeved lab coat with knit cuffs and over-the-cuff rubber

gloves. Use protective eyewear. Wear a protective mask if
appropriate.

Wipe-Down

• Open the door to its full open position.

• Wipe down the interior surfaces of the cabinet with 70% ethanol, or a

suitable disinfectant, and allow to dry.
CAUTION: Do not use bleach on the stainless steel liner. Use
disinfectants compatible with stainless steel.

Loading Evidence

• Load the evidence so that air flows around all surfaces of it. Do not

overload the cabinet. Do not place objects directly against the side
walls.

• Do not obstruct the intake prefilter (bottom of the door) or the exhaust

air discharge, on top of the cabinet.

• Large objects should not be placed close together.

• Slowly close the door and use a padlock (not supplied), or secure it

with a chain of custody seal.