Chapter 6: Maintaining the Cabinet

Product Service 1-800-522-7658

18

Service Operations

Door Evidence Prefilter Replacement:

1. Pull the prefilter frame straight away from the door, as shown in

Figure 6-1.

2. Wearing appropriate gloves, and working on the inside surface of the

prefilter frame, carefully pull out the Evidence prefilter. The Evidence
prefilter can now be examined for trace evidence, or stored with the
evidence in a sealed bag for future reference.

3. Examine the door prefilter. If it is contaminated from the evidence drying

operation, or from dust, replace it with a new one, such that the orange
side of the prefilter faces you (away from the front of the prefilter frame).

4. Press the prefilter frame back into the door, engaging the fastener strips.

Door Prefilter Replacement:

1. Follow Steps 1, 3, and 4 above.

Figure 6-1