Fan/UV Light Switch

The cabinet Fan/UV Light switch is located on the upper front of the cabinet,
as shown in Figure 5-1. Turning the switch to the “Fan” position starts the
blower, drawing air into through the cabinet. Turning the switch to the “UV
On” position will allow the UV light to turn on when the door is closed and
the UV Security lock is in the “ON” position.

UV Security Lock

The UV security lock, located to the right of the Fan/UV Light switch, allows
the user to lockout the UV, preventing its operation. Before evidence is
loaded into the cabinet, it is recommended that the lock be turned to the
“OFF” position, and the key removed. This will prevent the accidental
activation of the UV light while evidence is drying. The key should be stored
in a convenient, secure location, known to all of the users to prevent its loss.
Recommended locations include hanging from the bar at the top of the liner,
or in the lower storage area.

Filter System Light

The Filter System Light will illuminate if the differential pressure across the
filters becomes excessive. If lit, first inspect or replace the two prefilters. If lit
with clean prefilters, replace the HEPA filter.

NOTE: If the Filter System Light illuminates during use, finish the drying of
the evidence in the cabinet, and then correct the problem before using the
cabinet again.