Labconco Protecto Series Evidence Drying Cabinets 3390001 User Manual
Page 27
Advertising
![background image](/manuals/738641/27/background.png)
Chapter 7: Troubleshooting
Product Service 1-800-522-7658
23
PROBLEM
CAUSE
CORRECTIVE ACTION
Filter System Light
turns on
Blockage of the
intake prefilters
Check the inlet prefilters; replace if
needed. Ensure that there is no
obstruction of the exhaust on top of the
cabinet.
HEPA filter loading
Replace the HEPA filter.