Chapter 5: Using The Cabinet

Product Service 1-800-522-7658

16

Unloading Evidence

• Wear a long sleeved lab coat with knit cuffs and over-the-cuff rubber

gloves. Use protective eyewear. Wear a protective mask if
appropriate.

• Carefully remove objects in the cabinet. Evidence that was considered

biohazardous when loaded should be handled as biohazardous. If your
protocol required the use of evidence prefilters, remove them now and
store them appropriately.

Wipe-Down

• Wear a long sleeved lab coat with knit cuffs and over-the-cuff rubber

gloves. Use protective eyewear. Wear a protective mask if
appropriate.

• Wipe down the interior surfaces of the cabinet with 70% ethanol, or a

suitable disinfectant, and allow to dry.

• Inspect and clean the prefilters at the bottom of the door, and at the

top of the liner.

• Dispose of rubber gloves appropriately, and have lab coat laundered

properly.

• Wash hands and arms thoroughly with germicidal soap.

Shutdown

• Close the door. If UV disinfection is desired, insert the UV Security

Lock Key, and turn it to the “ON” position. Turn the Fan/UV Light
switch to “UV”. The UV light will light.

• Turn the UV Light off after an appropriate disinfection time, typically

1-8 hours.

Shutdown

Close the door and push the switch to the center “Off” position.