Labconco Protector Storage Cabinets 99071 User Manual

Page 24

Advertising
background image

20

Product Service 1-800-522-7658, International 816-333-8811

Chapter 3: Getting Started

Vent Installation – Solvent Storage
Cabinet


To ventilate the cabinet properly remove both the
exterior and interior barrel plugs from the vent
connection that you are venting.

The lower vent connection should be used for heavier
than air vapors and the higher vent connection on the
back wall should be used for lighter than air vapors to
obtain the best fume removal within the cabinet.

The cabinet vent connections are designed to accept 2"
diameter threaded metal pipe material, which can be
purchased locally. Once the connection on the back of
the cabinet has been completed, remove both the
internal and external plugs from the remaining vent
connection to allow for proper airflow from the cabinet.

CAUTION: Retain the barrel plugs
used with your vent connections
should it become necessary to isolate
the cabinet from the vent connection
as ducted.