Chapter 2: Prerequisites

Product Service 1-800-522-7658

6

Location Requirements

The fume hood should be located away from traffic patterns,
doors, windows, fans, ventilation registers, and any other air-
handling device that could disrupt its airflow patterns. All
windows in the room should be closed.

Support Requirements

DO NOT install the fume hood on a cart, dolly, or mobile
bench. ALL Protector Hood installations must be permanent
and stationary. The supporting structure usually consists of a
base cabinet or stand and chemically-resistant work surface.

Exhaust Requirements

The exhaust duct connection has been designed for 12" nominal duct (12.75" OD)
to allow for minimum static pressure loss while operating at 100 fpm face
velocities. The 12" diameter exhaust duct also allows for proper transport
velocities away from the hood in the 1000 fpm to 2500 fpm range. The proper
exhaust volume and static pressure loss are listed for each hood model:

Latched Sash Full Open 28"

Hood

Size

Standard Model Description

Face Velocity

Exhaust

Volume

Static

Pressure Loss

4'

48" Protector XLE Hood

60 fpm

470 CFM

0.07" H

2

O

80 fpm

630 CFM

0.12" H

2

O

100 fpm

790 CFM

0.19" H

2

O

5'

60" Protector XLE Hood

60 fpm

620 CFM

0.10" H

2

O

80 fpm

830 CFM

0.18" H

2

O

100 fpm

1040 CFM

0.28" H

2

O

6'

72" Protector XLE Hood

60 fpm

770 CFM

0.12" H

2

O

80 fpm

1020 CFM

0.21" H

2

O

100 fpm

1280 CFM

0.33" H

2

O

7'

84" Protector XLE Hood

60 fpm

920 CFM

0.07" H

2

O

80 fpm

1220 CFM

0.12" H

2

O

100 fpm

1530 CFM

0.19" H

2

O

8’

96" Protector XLEHood

60 fpm

1070 CFM

0.08" H

2

O

80 fpm

1420 CFM

0.14" H

2

O

100 fpm

1780 CFM

0.22" H

2

O

!

!