Chapter 2: Prerequisites

Product Service 1-800-522-7658

7

Self-Closing 60% Sash Opening 18"

Hood

Size

Standard Model Description

Face Velocity

Exhaust

Volume

Static

Pressure Loss

4'

48" Protector XLE Hood

60 fpm

310 CFM

0.03" H

2

O

80 fpm

420 CFM

0.06" H

2

O

100 fpm

520 CFM

0.09" H

2

O

5'

60" Protector XLE Hood

60 fpm

410 CFM

0.04" H

2

O

80 fpm

550 CFM

0.08" H

2

O

100 fpm

690 CFM

0.12" H

2

O

6'

72" Protector XLE Hood

60 fpm

510 CFM

0.06" H

2

O

80 fpm

680 CFM

0.10" H

2

O

100 fpm

850 CFM

0.15" H

2

O

7'

84" Protector XLE Hood

60 fpm

610 CFM

0.03" H

2

O

80 fpm

820 CFM

0.06" H

2

O

100 fpm

1020 CFM

0.09" H

2

O

8’

96" Protector XLE Hood

60 fpm

710 CFM

0.04" H

2

O

80 fpm

940 CFM

0.06" H

2

O

100 fpm

1180 CFM

0.10" H

2

O

Proper blower selection can be determined from these exhaust requirements and
the total system static pressure loss. Contact Labconco Customer Service for
assistance in sizing a blower system.

Electrical Requirements

The Protector Hood models feature internal wiring for the fluorescent light
assembly and light switch. All internal wiring is terminated at the single point
wiring junction box for hook-up by a qualified electrician. The blower switch,
and light switch wires are also terminated at the single point wiring junction box
for hook-up by a qualified electrician. Refer to Chapter 3: Getting Started and
Appendix C: Protector XLE Specifications for the wiring diagram for proper
electrical installation.

Service Line Requirements

All service lines to the laboratory fume hood should be a minimum of 1/4"
outside diameter (3/8" diameter for a cold water gooseneck), copper (brass for
natural gas), and equipped with an easily accessible shut-off valve, should
disconnection be required. If the service line pressure exceeds 40 PSI, it must be
equipped with a pressure regulator to reduce the line pressure. Please check with
local codes for other requirements.

Space Requirements

The dimensions for the different models are shown in Appendix B: Protector XLE
Dimensions
.

Your Next Step

After you have determined that the location you have selected accommodates the
installation and operational requirements of your fume hood, you are ready to
begin installation. Proceed to Chapter 3: Getting Started.