Environmental Conditions

• Indoor use only.

• Maximum altitude: 10,000 feet (3,048 meters).

• Ambient temperature range: 41° to 104°F (5° to 40°C).

• Maximum relative humidity: 80% for temperatures up to 88°F (31°C),

decreasing linearly to 50% relative humidity at 104°F (40°C).

• Main supply voltage fluctuations not to exceed ±10% of the nominal

voltage.

• Transient over-voltages according to Installation Categories II (Over-

voltage Categories per IEC 1010). Temporary voltage spikes on the AC
input line that may be as high as 1500V for 115V models and 2500V for
230V models are allowed.

• Used in an environment of Pollution degrees 2 (i.e., where normally only

non-conductive atmospheres are present). Occasionally, however, a
temporary conductivity caused by condensation must be expected, in
accordance with IEC 664.